Important Information
Warranty
The SCXI-1190, SCXI-1191, and SCXI-1192 are warranted against defects in materials and workmanship for a period of one year from the
date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that
proves to be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects
in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National
Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives
notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be
uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before
any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are
covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical
accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent
editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected.
In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMER’S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR
DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY
THEREOF. This limitation of the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including
negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments
shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover
damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or
maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire,
flood, accident, actions of third parties, or other events outside reasonable control.
Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying,
recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National
Instruments Corporation.
Trademarks
CVI™, LabVIEW™, National Instruments™, ni.com™, NI-DAQ™, PXI™, SCXI™, and VirtualBench™ are trademarks of National Instruments
Corporation.
Product and company names mentioned herein are trademarks or trade names of their respective companies.
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF
RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN
ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT
INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE
IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY,
COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS
AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND
HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL
DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR
MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE
HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD
CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD
NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID
DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO
PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS.
BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING
PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN
COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL
INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING
THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE
INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN,
PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.
Compliance
FCC/Canada Radio Frequency Interference Compliance*
Determining FCC Class
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC
places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial locations only)
or Class B (for use in residential or commercial locations). Depending on where it is operated, this product could be subject to
restrictions in the FCC rules. (In Canada, the Department of Communications (DOC), of Industry Canada, regulates wireless
interference in much the same way.)
Digital electronics emit weak signals during normal operation that can affect radio, television, or other wireless products. By
examining the product you purchased, you can determine the FCC Class and therefore which of the two FCC/DOC Warnings
apply in the following sections. (Some products may not be labeled at all for FCC; if so, the reader should then assume these are
Class A devices.)
FCC Class A products only display a simple warning statement of one paragraph in length regarding interference and undesired
operation. Most of our products are FCC Class A. The FCC rules have restrictions regarding the locations where FCC Class A
products can be operated.
FCC Class B products display either a FCC ID code, starting with the letters EXN,
or the FCC Class B compliance mark that appears as shown here on the right.
FCC/DOC Warnings
This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance with the instructions
in this manual and the CE Mark Declaration of Conformity**, may cause interference to radio and television reception.
Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department
of Communications (DOC).
Changes or modifications not expressly approved by National Instruments could void the user’s authority to operate the
equipment under the FCC Rules.
Class A
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC
Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and
used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct
the interference at his own expense.
Canadian Department of Communications
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
Class B
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the
FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can
be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
•
•
•
•
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Canadian Department of Communications
This Class B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe B respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
European Union - Compliance to EEC Directives
Readers in the EU/EEC/EEA must refer to the Manufacturer's Declaration of Conformity (DoC) for information** pertaining
to the CE Mark compliance scheme. The Manufacturer includes a DoC for most every hardware product except for those bought
for OEMs, if also available from an original manufacturer that also markets in the EU, or where compliance is not required as
for electrically benign apparatus or cables.
*
Certain exemptions may apply in the USA, see FCC Rules §15.103 Exempted devices, and §15.105(c). Also available in
sections of CFR 47.
** The CE Mark Declaration of Conformity will contain important supplementary information and instructions for the user or
installer.
Conventions
The following conventions are used in this manual:
»
The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to
pull down the File menu, select the Page Setup item, and select Options
from the last dialog box.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to
avoid injury, data loss, or a system crash.
This icon denotes a warning, which advises you of precautions to take to
avoid being electrically shocked.
bold
Bold text denotes items that you must select or click on in the software,
such as menu items and dialog box options. Bold text also denotes
parameter names.
italic
Italic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. This font also denotes text that is a placeholder for a word
or value that you must supply.
monospace
Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples.
This font is also used for the proper names of disk drives, paths, directories,
programs, subprograms, subroutines, device names, functions, operations,
variables, filenames and extensions, and code excerpts.
monospace bold
Bold text in this font denotes the messages and responses that the computer
automatically prints to the screen. This font also emphasizes lines of code
that are different from the other examples.
SCXI-119X
Refers to the SCXI-1190 module, the SCXI-1191 module, and the
SCXI-1192 module.
Contents
Chapter 1
About the SCXI-119X....................................................................................................1-1
Installing the Software...................................................................................................1-2
Safety Instructions .........................................................................................................1-3
Installing the Hardware..................................................................................................1-3
Configuring and Self-Testing the SCXI-119X...............................................................1-9
Manually Adding Modules..............................................................................1-10
Self-Test Verification......................................................................................1-11
Removing the SCXI-119X.............................................................................................1-13
Chapter 2
Functional Overview......................................................................................................2-1
RF Switching Precautions..............................................................................................2-3
Connecting Signals ........................................................................................................2-5
Multiplexing to RF Sources.............................................................................2-6
Expanding the RF Multiplexer........................................................................2-8
Appendix A
Specifications
Appendix B
Common Questions
© National Instruments Corporation
vii
SCXI-1190/1191/1192 User Manual
Contents
Appendix C
Technical Support Resources
Glossary
Index
SCXI-1190/1191/1192 User Manual
viii
ni.com
1
Installing and Configuring the
SCXI-119X
This manual describes the electrical and mechanical characteristics of the
SCXI-1190 (quad 1 × 4, 50 Ω, 1.3 GHz bandwidth) and SCXI-1191 (quad
1 × 4, 50 Ω, 4 GHz bandwidth) multiplexer modules, and the SCXI-1192
(8-channel, SPDT, 50 Ω, 18 GHz bandwidth) relay module for the SCXI
bus. It also contains information concerning their installation, operation,
and safety instructions.
About the SCXI-119X
This section summarizes the features and operation of the SCXI-119X
switch modules. In addition, refer to Appendix A, Specifications,
for detailed specifications of the switch modules.
The SCXI-1190/1191 modules are general-purpose, quad 4-channel,
high-bandwidth multiplexing switches. The SCXI-1192 module is a
general-purpose, 8-channel, SPDT, high-bandwidth relay module.
The SCXI-119X modules use single-pole double-throw, high-bandwidth
relays capable of switching signals from DC to 1.3 GHz bandwidth
(SCXI-1190), from DC to 4 GHz bandwidth (SCXI-1191), and from DC to
18 GHz bandwidth (SCXI-1192). The characteristic impedance of the
channels is 50 Ω.
and the maximum rated current is 1 A at 24 V. For the SCXI-1191, the
maximum rated voltage of the switch is 30 V, and the maximum rated
current is 0.33 A at 30 V. For the SCXI-1192, the maximum rated voltage
of the switch is 30 V, and the maximum rated current is 2 A at 30 V.
See Appendix A, Specifications, for full specifications.
© National Instruments Corporation
1-1
SCXI-1190/1191/1192 User Manual
Chapter 1
Installing and Configuring the SCXI-119X
Installing the Software
You can control the SCXI-119X programmatically in an application
development environment (ADE) using NI-SWITCH. The supported
ADEs include LabVIEW, LabWindows/CVI, Visual Basic, and C or C++
environments. To install software for the SCXI-119X, complete the
following:
1. Insert your NI-SWITCH software CD into your CD-ROM drive.
The installation window should appear automatically.
Note If the installation window does not appear, double-click the My Computer icon on
your PC desktop. Find the CD drive, and double-click it. In the CD directory, double-click
setup.exe.
The NI-SWITCH and NI-DMM installation window offers three
choices:
•
•
•
Install NI-DMM—Choose this selection if you want to install
only the DMM software.
Install NI-SWITCH—Choose this selection if you want to install
only the switches software.
Install NI-SWITCH and NI-DMM—Choose this selection if
you want to install both the DMM and switches software.
2. Click the Install NI-SWITCH option.
3. To install the instrument driver, VirtualBench, and application
development examples, choose Programmatic and Interactive
Support. To install only the instrument driver files for VirtualBench,
choose Interactive Support Only.
Note Choosing Interactive Support Only does not allow you to program the instrument
with any programming languages.
4. When installing NI-SWITCH, look at the Development
Environments panel.
•
If you click the Advanced button on this panel, you can custom
install National Instruments drivers such as NI-DAQ, NI-IVI, and
NI-VISA.
uncheck the driver(s) you do not want installed.
If a driver is already unchecked, your computer has the same driver or
a newer version of the driver already installed.
SCXI-1190/1191/1192 User Manual
1-2
ni.com
Chapter 1
Installing and Configuring the SCXI-119X
Note If a newer version of a driver is present on your PC, the installer will not overwrite
the driver.
Safety Instructions
Cautions Do not operate damaged equipment. The safety protection features built into this
instrument can become impaired if the instrument becomes damaged in any way. If the
instrument is damaged, do not use it until service-trained personnel can check its safety.
If necessary, return the instrument to National Instruments for service and repair to ensure
that its safety is not compromised.
Do not operate this instrument in a manner that contradicts the information specified in this
document. Misuse of this instrument could result in a shock hazard.
Do not substitute parts or modify equipment. Because of the danger of introducing
additional hazards, do not install unauthorized parts or modify the instrument. Return the
instrument to National Instruments for service and repair to ensure that its safety is not
compromised.
Connections that exceed any of the maximum signal ratings on the SCXI-119X can create
a shock or fire hazard or can damage any or all of the devices connected to the SCXI-119X.
National Instruments is not liable for any damages or injuries resulting from incorrect
signal connections.
Clean the instrument and accessories by brushing off light dust with a soft, nonmetallic
brush. Remove other contaminants with a stiff nonmetallic brush. The unit must be
completely dry and free from contaminants before returning to service.
Installing the Hardware
The following section describes how to unpack, install, and connect your
SCXI-119X for use with an SCXI chassis.
Caution Make sure you read and understand all the safety information in this manual
before you install an SCXI module.
© National Instruments Corporation
1-3
SCXI-1190/1191/1192 User Manual
Chapter 1
Installing and Configuring the SCXI-119X
Unpacking
Your SCXI-119X module is shipped in an antistatic package to prevent
electrostatic damage to the module. Electrostatic discharge can damage
several components on the module. To avoid such damage in handling the
module, take the following precautions:
•
Ground yourself using a grounding strap or by holding a grounded
object.
•
Touch the antistatic package to a metal part of your computer chassis
before removing the module from the package.
Caution Never touch the exposed pins of connectors.
Remove the module from the package and inspect the module for loose
components or any sign of damage. Notify National Instruments if the
module appears damaged in any way. Do not install a damaged module into
your computer.
Store your SCXI-119X module in the antistatic envelope when not in use.
Installing the SCXI-119X Module into an SCXI Chassis
You need the following items to complete the installation:
•
•
SCXI-119X module
SCXI chassis
SCXI-1190/1191/1192 User Manual
1-4
ni.com
Chapter 1
Installing and Configuring the SCXI-119X
Follow these steps to install your SCXI-119X module into an SCXI chassis
while referring to Figure 1-1 and your chassis documentation:
5
6
®
7
S
C
M
X
M
A
I
E
IN
F
R
A
8
2
1
4
3
1
2
3
4
Slot 1
Slot 2
Slot 3
Slot 4
5
6
7
8
SCXI Chassis (SCXI-1000 Shown)
SCXI Chassis Power Switch
SCXI-119X (SCXI-1190 Shown)
SCXI Module Thumbscrews
Figure 1-1. Installing the SCXI-119X into an SCXI Chassis
1. Turn off the computer that contains the DAQ device or computer-based
instrument, or disconnect the computer from your SCXI chassis.
2. Turn off the SCXI chassis.
Caution Do not insert the SCXI-119X module into a chassis that is powered on because
doing so could damage the module.
3. Remove the protective plastic caps from the thumbscrews of the
module so that the module will insert correctly.
4. Insert the SCXI-119X module into slot 1. Gently guide the module into
the module guides and push it to the back of the slot until the
connectors make good contact. The module must be firmly engaged;
however, do not force the module into place. If the module is not
sliding easily into the chassis, check for correct insertion into the
module guides.
© National Instruments Corporation
1-5
SCXI-1190/1191/1192 User Manual
Chapter 1
Installing and Configuring the SCXI-119X
Note When installing the SCXI-119X module in an SCXI chassis, fill slot 1 first, then fill
slot 2 with the next SCXI-119X.
5. Insert any other SCXI modules into the remaining slots in the same
manner as described in step 4.
6. Secure all the SCXI-119X modules to the SCXI chassis by tightening
both module thumbscrews.
Connecting the SCXI-119X to a DAQ Device or Computer-Based
Instrument
You need the following items:
•
SCXI chassis with the SCXI modules installed (Refer to the Installing
the SCXI-119X Module into an SCXI Chassis section for more
information.)
•
•
SCXI cable assembly, which consists of a cable adapter and a cable
An installed National Instruments DAQ device or computer-based
instrument, such as the NI 4060, that is able to control the SCXI
chassis
•
1/4 in. flathead screwdriver
Consult your SCXI chassis documentation, other SCXI module
documentation, and DAQ device or computer-based instrument
documentation for additional instructions and warnings. Your SCXI-119X
module and any other SCXI modules should already be installed in the
chassis according to their installation instructions.
You must select one of the modules in the chassis as the cabled
module—the module to which the SCXI cable assembly connects. If the
chassis contains at least one SCXI-1140 module, select an SCXI-1140 as
the cabled module; otherwise, select one of the remaining modules.
Note The SCXI-119X modules operate only in multiplexed mode.
SCXI-1190/1191/1192 User Manual
1-6
ni.com
Chapter 1
Installing and Configuring the SCXI-119X
To connect the chassis to a DAQ device, complete the following steps while
referring to Figure 1-2:
1
4
3
®
2
1
2
SCXI Chassis Power Switch
Cable Adapter
3
4
Cable
DAQ Device or Computer-Based Instrument
Figure 1-2. Connecting an SCXI Chassis to a DAQ Device
or Computer-Based Instrument
1. Power off both the SCXI chassis and the computer that contains the
DAQ device or computer-based instrument.
2. Insert the cable adapter into the back of the SCXI chassis aligned with
the module that is to be connected to the DAQ device or
computer-based instrument. See your cable assembly installation
guide for more specific information.
3. Connect the cable to the back of the cable adapter, ensuring that the
cable fits securely.
4. Connect the other end of the cable to the DAQ device or
computer-based instrument that you will use to control the SCXI
system.
© National Instruments Corporation
1-7
SCXI-1190/1191/1192 User Manual
Chapter 1
Installing and Configuring the SCXI-119X
5. Check the cable installation, making sure the connectors are securely
fastened at both ends.
6. Power on the SCXI chassis.
7. Power on the computer.
8. If you have already installed the appropriate software, you are ready to
configure the SCXI-119X. If you have not installed the appropriate
software, refer to the Installing the Software section.
Connecting the SCXI-119X to a DAQ Device in a PXI
Combination Chassis
To connect the SCXI-119X to a DAQ device in a PXI combination chassis,
such as the PXI-1010 or PXI-1011, you need the following items:
•
•
PXI combination chassis with the SCXI modules installed
National Instruments PXI MIO DAQ device, or a computer-based
instrument that is capable of controlling SCXI, installed in the right
most slot of the PXI section of the combination chassis
A PXI combination chassis can use both SCXI and PXI modules at the
same time. Because the connection between the SCXI backplane and the
PXI backplane is made through a local bus interface to a PXI multifunction
I/O module installed in the right most slot, no external cabling is needed to
connect one PXI multifunction I/O module to the SCXI bus.
You should have already installed your SCXI-119X module and any other
SCXI modules in the chassis according to their installation instructions.
To use the SCXI-119X module with a DAQ device or computer-based
instrument in a PXI combination chassis, turn on the SCXI chassis.
documentation, and DAQ device or computer-based instrument
documentation for additional instructions and warnings.
If you have already installed the appropriate software, you are ready to
configure the SCXI-119X. If you have not installed the appropriate
software, refer to the Installing the Software section.
SCXI-1190/1191/1192 User Manual
1-8
ni.com
Chapter 1
Installing and Configuring the SCXI-119X
Configuring and Self-Testing the SCXI-119X
Run Measurement & Automation Explorer (MAX) to configure and test
your SCXI-119X. If you need help during the configuration process, open
the Measurement & Automation Help file by selecting Help»Help Topics.
1. Double-click the Measurement & Automation Explorer icon on
your desktop.
2. To add a new chassis, right-click Devices and Interfaces and select
Create New.
3. Select the appropriate chassis from the list box and click Finish.
4. To configure the chassis, select the Chassis ID. The chassis ID is an
arbitrary number that you choose for use by the software to uniquely
specify the chassis.
5. Select the Chassis Address (refer to your SCXI Chassis User Manual
for further information).
For the SCXI-2000, you also need to select the Baud Rate and
COM Port.
You now have the choice of auto-detecting your modules or manually
adding them depending on your application.
Auto-Detecting Modules
If you selected auto-detect and the chassis is not an SCXI-2000 remote
chassis, you must have your chassis connected to your National
Instruments device.
Note Make sure you have your chassis powered on.
1. Click Next.
2. Select your communication path, except in the case of the SCXI-2000,
and click Next.
3. Select the module cabled to your communication path, except in the
case of the SCXI-2000 or if your module supports Auto Cable Detect,
and click Finish.
© National Instruments Corporation
1-9
SCXI-1190/1191/1192 User Manual
Chapter 1
Installing and Configuring the SCXI-119X
Your SCXI chassis and SCXI-119X modules are now configured in the
software with your modules. To configure the properties for a particular
module, perform the following steps:
1. Expand Devices and Interfaces. You can see the chassis selected in
the list.
2. Expand Chassis.
3. Right-click the module you want to configure and select Properties.
Tabs are displayed containing attributes for different properties of the
module. You can select any tab to see and/or change any attributes.
Manually Adding Modules
If you selected not to auto-detect your SCXI modules, you must add each
of your modules separately.
1. Click Finish in the SCXI Auto-Detect window.
2. Expand Devices and Interfaces. You will see the chassis you selected
in the list.
3. Expand Chassis.
4. Right-click the slot that is cabled to your device and select Insert. For
DAQ devices this slot is the slot cabled to the DAQ device. Select the
module that is intended for that slot and click Next.
5. You are now given the option to make this the module that controls the
chassis. Select the DAQ device or computer-based instrument you
cabled to the chassis using the Connected to: control. Confirm that
there is a check in the checkbox labeled This device will control the
chassis and click Next.
6. You are presented with a series of properties for the module. Select the
appropriate values for the attributes and click Next. After making your
selections, click Finish.
You are now finished configuring the slot. Configure the rest of your
modules in the same manner.
To remove a module or chassis, right-click on the chassis or module you
want to remove and select Delete.
To test the chassis configuration, right-click on the chassis and select Test.
SCXI-1190/1191/1192 User Manual
1-10
ni.com
Chapter 1
Installing and Configuring the SCXI-119X
Self-Test Verification
To test the successful configuration of your system, complete the following
steps:
1. Double-click the Measurement & Automation Explorer icon on
your PC desktop.
2. Verify that the chassis power is on and that the chassis is correctly
connected to a DAQ device or computer-based instrument.
3. Expand Devices and Interfaces.
4. From the list of devices and interfaces that appears, locate the chassis
you want to test. Right-click on the chassis and select Test. If the
communication test is successful, a message The chassis has
been verifiedappears.
5. Click OK.
Note If the communication test is unsuccessful, refer to the Troubleshooting Self-Test
Verification section.
Your SCXI system should now operate properly with your ADE software.
If you experience problems between your SCXI system and your ADE,
refer to the Troubleshooting Self-Test Verification section. Additionally,
you may find it helpful to refer to your NI-SWITCH documentation,
your ADE software documentation, and your chassis documentation.
National Instruments documentation is available to download free at
ni.com/manuals
Troubleshooting Self-Test Verification
If the preceding test did not verify your chassis configuration, take the
following steps to successfully complete system configuration:
•
If you get the warning message Unable to test chassis at this
time, verify that you have designated a module as being connected to
a DAQ device or computer-based instrument. Take the following
troubleshooting actions:
a. Double-click the Measurement & Automation Explorer icon
on your PC desktop.
b. Expand Devices and Interfaces.
d. Right-click the module that is connected to your DAQ device or
computer-based instrument and select Properties.
© National Instruments Corporation
1-11
SCXI-1190/1191/1192 User Manual
Chapter 1
Installing and Configuring the SCXI-119X
e. Change the Connected to selection from None to Connected to:
Device x.
•
If you get the warning message Failedto findfollowed by the
module codes and the message Unable to communicate with
chassis, take the following troubleshooting actions:
a. Make sure that the SCXI chassis is powered on.
b. Make sure the cable between the SCXI chassis and DAQ device or
computer-based instrument is properly connected.
c. Inspect the cable and the connectors for any bent pins.
d. Make sure you are using the correct National Instruments cable
assembly.
e. Test your DAQ device or computer-based instrument to verify it is
working properly. Refer to your DAQ device or computer-based
•
If you get the warning message Failed to find, followed by
module codes and then the message Instead found: module with
ID 0Xxx:
module is in the specified slot.
b. Delete the incorrect module as described in the Removing the
SCXI-119X from an SCXI Chassis section.
c. Add the correct module as described in the Manually Adding
•
If you get the warning message Failed to find, followed by a
module code and then the message Slot x is empty:
a. Check to see if the configured module is installed in the specified
slot. If not, install the module by referring to the Installing the
Hardware section.
and remove the module as specified in the Removing the
SCXI-119X from an SCXI Chassis section. Verify that no
connector pins are bent on the rear signal connector. Reinstall the
module, ensuring the module is fully inserted and properly aligned
in the slot.
section and retest your SCXI chassis.
SCXI-1190/1191/1192 User Manual
1-12
ni.com
Chapter 1
Installing and Configuring the SCXI-119X
Removing the SCXI-119X
This section describes how to remove the SCXI-119X from an SCXI
chassis.
Removing the SCXI-119X from an SCXI Chassis
Make sure you read and understand all the safety information in this
manual before you remove an SCXI module.
Consult the documentation for your SCXI or PXI (if you are using a
combination chassis) chassis and accessories for additional instructions
and warnings.
To remove the SCXI-119X module from an SCXI chassis, complete the
following steps while referring to Figure 1-3:
2
1
®
3
4
1
2
SCXI Chassis Power Switch
SCXI Chassis (SCXI-1000 Shown)
3
4
SCXI-119X (SCXI-1190 Shown)
SCXI Module Thumbscrews
Figure 1-3. Removing the SCXI-119X
© National Instruments Corporation
1-13
SCXI-1190/1191/1192 User Manual
Chapter 1
Installing and Configuring the SCXI-119X
1. Turn off the SCXI chassis power.
2. Remove all signals connected to the SCXI-119X.
Caution Do not remove the SCXI-119X module from a chassis that is turned on because
doing so could damage the module.
3. Rotate the thumbscrews securing the SCXI-119X to the chassis
counter-clockwise until they are loose, but do not completely remove
the thumbscrews.
4. Remove the SCXI-119X by pulling steadily on both thumbscrews until
the module slides completely out.
Removing the SCXI-119X from Measurement & Automation Explorer
To remove a module from MAX, complete the following steps:
1. Double-click the Measurement & Automation Explorer icon on
your PC desktop.
2. Expand Devices and Interfaces. Locate the chassis in the list of
installed devices.
3. Expand Chassis.
4. Right-click on the module or chassis you want to delete and select
Delete.
5. You are presented a confirmation window. Click Yes to continue
deleting the module or chassis or click No to cancel this action.
Note Deleting an SCXI chassis deletes all modules in the chassis. All configuration
information for these modules is also lost.
Your SCXI chassis and/or SCXI module(s) should now be removed from
the list of installed devices in MAX.
SCXI-1190/1191/1192 User Manual
1-14
ni.com
2
Using the SCXI-119X
This chapter contains a functional overview and connection possibilities of
the SCXI-119X modules.
Functional Overview
The SCXI-1190 uses single-pole double-throw, high-bandwidth relays
capable of switching signals from DC to 1.3 GHz. Functionally identical,
the SCXI-1191 has a bandwidth of DC to 4 GHz. The SCXI-1192 module
is a general-purpose, 8-channel, high-bandwidth relay module; it uses
single-pole double-throw, high-bandwidth relays capable of switching
signals from DC to 18 GHz.
ni.com© National Instruments Corporation
2-1
SCXI-1190/1191/1192 User Manual
Chapter 2
Using the SCXI-119X
Figure 2-1 illustrates the key functional components of the
SCXI-1190/1191 multiplexer modules.
Rear Signal
Connector
SCXIbus
Relay Controller
and
Drivers
Relay Control Bus
comA 3A
2A
1A 0A
comB 3B
2B
1B 0B
comC 3C 2C 1C 0C
comD 3D 2D 1D 0D
Figure 2-1. SCXI-1190/1191 Modules Block Diagram
SCXI-1190/1191/1192 User Manual
2-2
ni.com
Chapter 2
Using the SCXI-119X
Figure 2-2 illustrates the key functional components of the SCXI-1192
relay module.
Rear Signal
Connector
SCXIbus
Relay Controller
com0 NC0 NO0 com1 NC1 NO1 com2 NC2 NO2 com3 NC3 NO3 com4 NC4 NO4 com5 NC5 NO5 com6 NC6 NO6 com7 NC7 NO7
Figure 2-2. SCXI-1192 Module Block Diagram
RF Switching Precautions
The SCXI-119X are reflective switches, meaning that any channels not
switched to the common channel are unterminated, and any signal on an
unterminated channel is reflected to its source. For most low-power
switching applications, this reflection is not a problem. However, operation
with an unterminated output can damage some high-power RF sources.
Consult your RF source documentation for more information about
connecting to unterminated channels.
Warning Do not exceed the channel-to-ground rating of 24 Vrms or DC on the SCXI-1190,
that exceed the maximum voltage for the SCXI-119X can result in an electrical shock
hazard and damage to the switch module and any or all of the modules connected to the
© National Instruments Corporation
2-3
SCXI-1190/1191/1192 User Manual
Chapter 2
Using the SCXI-119X
SCXI backplane. National Instruments is not liable for any damages or injuries resulting
from exceeding maximum voltage ratings. Refer to Appendix A, Specifications, for
specific information.
Caution Always disconnect all signals from the front connectors and power off the SCXI
chassis before removing the SCXI-119X from the chassis. Observe electrostatic discharge
handling precautions during and after removal of the module, or when connecting and
disconnecting signals on the front panel.
Contact Protection
Caution The contacts of the relays are rated for a life of 5 × 106 mechanical operations
when operated with no signal or a very small signal present. When high-power loads are
switched, arcing can occur across the relay contacts that will reduce their useful life. See
Appendix A, Specifications, for lifetime ratings for each module. If your setup allows, it is
always best to turn off the input signals connected to the switch module during the
switching operation to avoid contact arcing within the switch. Switching large inductive
loads can also cause contact arcing. In such instances, install a flyback diode or varistor
across the inductance to protect the switch.
SCXI-1190/1191/1192 User Manual
2-4
ni.com
Chapter 2
Using the SCXI-119X
Connecting Signals
The SCXI-1190 has push-on SMB connectors, and the SCXI-1191/1192
have screw-on SMA connectors as shown in Figure 2-3.
NATIONAL
NATIONAL
NATIONAL
INSTRUMENTS
INSTRUMENTS
INSTRUMENTS
Figure 2-3. SCXI-1190 SMB Connectors, SCXI-1191/1192 SMA Connectors
Note The power-up and reset state of the different multiplexer banks on the
SCXI-1190/1191 are as follows: banks A and C, channel 0 is connected to the COM
terminal; banks B and D, channel 3 is connected is connected to the COM terminal. The
reset mode of the SCXI-1192 connects NC to COM for all relays. Because it uses latching
relays, the SCXI-1192 does not have a specific power-up state. Its relays will retain their
state if they lose power.
© National Instruments Corporation
2-5
SCXI-1190/1191/1192 User Manual
Chapter 2
Using the SCXI-119X
Multiplexing to RF Sources
Figure 2-4 shows how to use the SCXI-1190/1191 to select one of four
RF sources to be routed to a RF measurement device.
NATIONAL
INSTRUMENTS
RF Source
1
RF Source
2
RF Measurement Device
• Scope
• Spectrum Analyzer
• Power Meter
RF Source
3
RF Source
4
Figure 2-4. Connecting the SCXI-1190/1191
SCXI-1190/1191/1192 User Manual
2-6
ni.com
Chapter 2
Using the SCXI-119X
Figure 2-5 shows how to use the SCXI-1192 to route signals from two
RF sources to one RF measurement device.
NATIONAL
INSTRUMENTS
RF Source
1
RF Source
2
RF Measurement Device
• Scope
• Spectrum Analyzer
• Power Meter
Figure 2-5. Connecting the SCXI-1192
© National Instruments Corporation
2-7
SCXI-1190/1191/1192 User Manual
Chapter 2
Using the SCXI-119X
Expanding the RF Multiplexer
Figure 2-6 shows how to expand the SCXI-1190/1191 by cascading stages
together. In the specific case of Figure 2-6, the module is used to create a
13-to-1 multiplexer with a two-stage architecture. You can cascade several
modules to provide even larger multiplexers.
Note Cascading multiplexers increases insertion loss, creates propagation delays, and can
have other RF effects.
3D
2D
comD MUX
1D
0D
D
3A
2A
1A
0A
3C
2C
1C
0C
comA MUX
A
comC MUX
C
3B
2B
1B
0B
comB MUX
B
Figure 2-6. Expanding the SCXI-1190/1191
SCXI-1190/1191/1192 User Manual
2-8
ni.com
Chapter 2
Using the SCXI-119X
Figure 2-7 shows one possible expansion using two SCXI-1192 modules.
In this case, the switches on two SCXI-1192 modules are interconnected to
create a 2 × 4 matrix. Multiple configurations are possible using one or
more SCXI-1192 modules.
Note Cascading switches increases insertion loss, creates propagations delays, and can
have other RF effects.
RF
RF
RF
RF
Source 1
Source 2
Source 3
Source 4
0B
1B
com B
Switch 1
Switch 2
Instrument
1
Switch 0
Switch 3
Switch 5
Switch 6
Instrument
2
Switch 4
Switch 7
Figure 2-7. Expanding the SCXI-1192
© National Instruments Corporation
2-9
SCXI-1190/1191/1192 User Manual
A
Specifications
This appendix lists the specifications for the SCXI-1190, the SCXI-1191,
and the SCXI-1192 multiplexer modules. These specifications are typical
at 25 °C unless otherwise noted.
SCXI-1190
Input Characteristics
Number of banks.................................... 4
Number of channels per bank ................ 4
Common-mode voltage
Channel to earth.............................. 24 Vrms or DC
Maximum switching voltage
AC................................................... 24 Vrms
DC................................................... 24 VDC
Maximum switching capacity per channel
AC................................................... 1 Arms at 24 Vrms
DC................................................... 1 ADC at 24 VDC
Maximum switching
power per channel .................................. 24 W
Contact on resistance (initial) ................ 100 mΩ max
Contact material ..................................... Gold-clad silver
© National Instruments Corporation
A-1
SCXI-1190/1191/1192 User Manual
Appendix A
Specifications for SCXI-1190
RF Performance Characteristics
Characteristic impedance (Z0) ................50 Ω
Insertion loss at:
≤100 MHz........................................<0.4 dB
≤500 MHz........................................<0.9 dB
≤1.3 GHz .........................................<1.5 dB
≤2 GHz ............................................<3 dB
VSWR at:
≤100 MHz........................................<1.15
≤500 MHz........................................<1.35
≤1.3 GHz .........................................<1.5
≤2 GHz ............................................<2.5
Isolation:
≤500 MHz........................................<–62 dB
≤1.3 GHz .........................................<–50 dB
≤2 GHz ............................................<–30 dB
Rise time.................................................<300 ps
Signal delay ............................................<3 ns
Maximum RF carry power
at 900 MHz .............................................10 W
Note Refer to the cautions and warning in Chapter 2, Using the SCXI-119X, for important
information about using your SCXI-1190 with high-power signals.
Dynamic Characteristics
Relay operate time (at 20 °C)
Typical.............................................15 ms
Relay release time (at 20 °C)
Typical.............................................15 ms
SCXI-1190/1191/1192 User Manual
A-2
ni.com
Appendix A
Specifications for SCXI-1190
Expected life
Mechanical (no load) ...................... 5 × 106 operations
Electrical at maximum
switching capacity........................... 105 operations
Caution Exceeding the maximum switching capability decreases the expected life of the
SCXI-1190.
Power Requirement
+5 VDC
Maximum (all relays closed) .......... 600 mA
Physical
Dimensions............................................. 17.30 by 19.56 by 3.02 cm
(6.81 by 7.70 by 1.19 in.)
I/O connectors........................................ 5 SMB male per bank;
4 banks
Environment
Operating temperature ..........................0 to 50 °C
Storage temperature..............................–20 to 70 °C
Relative humidity................................... 10% to 90% noncondensing
Electromagnetic Compatibility
EMC/EMI............................................... CE, C-Tick, and FCC Part 15
(Class A) Compliant
Electrical emissions................................ EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
Electrical immunity................................ Evaluated to EN 61326-1:1997
A1:1998, Table 1
Note For full EMC and EMI compliance, you must operate this device with shielded
cabling. See the Declaration of Conformity (DoC) for this product for any additional
regulatory compliance information. To obtain the DoC for this product, click
the DoCs by product family. Select the appropriate product family, followed by your
© National Instruments Corporation
A-3
SCXI-1190/1191/1192 User Manual
Appendix A
Specifications for SCXI-1190
product, and a link to the DoC (in Adobe Acrobat format) appears. Click the Acrobat icon
to download or read the DoC.
Safety
Designed in accordance with IEC 61010-1, UL 3111-1, and CAN/CSA
C22.2 No. 1010.1 for electrical measuring and test equipment.
Installation category II1
Pollution degree 2
Certifications and Compliances
CE Mark Compliance
See the Declaration of Conformity sheet accompanying product.
1
Category II refers to local-level power distribution, such as that provided by a standard wall outlet.
SCXI-1190/1191/1192 User Manual
A-4
ni.com
Appendix A
Specifications for SCXI-1191
SCXI-1191
Input Characteristics
Number of banks.................................... 4
Number of channels per bank ................ 4
Common-mode voltage
Channel to earth.............................. 30 Vrms or DC
Maximum switching voltage
AC................................................... 30 Vrms
DC................................................... 30 VDC
Maximum switching capacity per channel
AC................................................... 0.33 Arms at 30 Vrms
DC................................................... 0.33 ADC at 30 VDC
Maximum switching
power per channel .................................. 10 W
Contact on resistance (initial) ................ 200 mΩ max
Contact material ..................................... Gold
RF Performance Characteristics
Characteristic impedance (Z0)................ 50 Ω
Insertion loss at:
< 2.5 GHz........................................ < 0.6 dB
< 4 GHz........................................... < 0.9 dB
VSWR at:
< 2.5 GHz........................................ < 1.3
< 4 GHz........................................... < 1.5
© National Instruments Corporation
A-5
SCXI-1190/1191/1192 User Manual
Appendix A
Specifications for SCXI-1191
Isolation:
< 4 GHz ...........................................< –55 dB
Maximum RF carry power
at 900 MHz .............................................10 W
Note Refer to the warning and cautions in Chapter 2, Using the SCXI-119X, for important
information about using your SCXI-1191 with high-power signals.
Dynamic Characteristics
Relay operate time (at 20 °C)
Typical.............................................15 ms
Relay release time (at 20 °C)
Typical.............................................15 ms
Expected life
Mechanical (no load).......................5 × 106 operations
Electrical at maximum
switching capacity ...........................105 operations
Caution Exceeding the maximum switching capability decreases the expected life of the
SCXI-1191.
Power Requirement
+5 VDC
Maximum (all relays closed)...........950 mW
Physical
Dimensions .............................................17.30 by 19.56 by 3.02 cm
(6.81 by 7.70 by 1.19 in.)
I/O connector ..........................................5 SMA female per bank
4 banks
SCXI-1190/1191/1192 User Manual
A-6
ni.com
Appendix A
Specifications for SCXI-1191
Environment
Operating temperature ..........................0 to 50 °C
Storage temperature..............................–20 to 70 °C
Relative humidity................................... 10% to 90% noncondensing
Electromagnetic Compatibility
EMC/EMI............................................... CE, C-Tick, and FCC Part 15
(Class A) Compliant
Electrical emissions................................ EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
Electrical immunity................................ Evaluated to EN 61326-1:1997
A1:1998, Table 1
Note For full EMC and EMI compliance, you must operate this device with shielded
cabling. See the Declaration of Conformity (DoC) for this product for any additional
regulatory compliance information. To obtain the DoC for this product, click
the DoCs by product family. Select the appropriate product family, followed by your
product, and a link to the DoC (in Adobe Acrobat format) appears. Click the Acrobat icon
to download or read the DoC.
Safety
Designed in accordance with IEC 61010-1, UL 3111-1, and CAN/CSA
C22.2 No. 1010.1 for electrical measuring and test equipment.
Installation category II1
Pollution degree 2
Certifications and Compliances
CE Mark Compliance
See the Declaration of Conformity sheet accompanying product.
1
Category II refers to local-level power distribution, such as that provided by a standard wall outlet.
© National Instruments Corporation
A-7
SCXI-1190/1191/1192 User Manual
Appendix A
Specifications for SCXI-1192
SCXI-1192
Input Characteristics
Number of relays ....................................8
Number of channels per relay.................2
Common-mode voltage
Channel to earth...............................30 Vrms or DC
Maximum switching voltage
DC....................................................30 VDC
Maximum carry current per channel
DC....................................................2 A
Maximum continuous input power per channel (continuous wave)
1–3 GHz...........................................120 W
3–8 GHz...........................................80 W
8–12.4 GHz......................................60 W
12.4–18 GHz....................................50 W
Contact on resistance (initial).................100 mΩ max
Contact material......................................Gold plate
RF Performance Characteristics
Characteristic impedance (Z0) ................50 Ω
Insertion loss at:
≤4 GHz ............................................≤0.2 dB
4 to 8 GHz .......................................≤0.3 dB
8 to 12.4 GHz ..................................≤0.4 dB
12.4 to 18 GHz ................................≤0.5 dB
VSWR at:
1 to 4 GHz .......................................≤1.15
4 to 8 GHz .......................................≤1.25
SCXI-1190/1191/1192 User Manual
A-8
ni.com
Appendix A
Specifications for SCXI-1192
8 to 12.4 GHz.................................. ≤1.35
12.4 to 18 GHz................................ ≤1.5
Isolation:
≤1 GHz............................................ ≤–85 dB
4 to 8 GHz....................................... ≤–70 dB
8 to 12.4 GHz.................................. ≤–65 dB
12.4 to 18 GHz................................ ≤–60 dB
Note Refer to the cautions and warning in Chapter 2, Using the SCXI-119X, for important
information about using your SCXI-1192 with high-power signals.
Dynamic Characteristics
Relay operate time (at 20 °C)
Typical ............................................ 15 ms
Relay release time (at 20 °C)
Typical ............................................ 15 ms
Expected life
Mechanical (no load) ...................... 5 × 106 operations
Electrical at 5 W ............................. 5 × 106 operations (to 3 GHz,
50 Ω, VSWR max 1.2)
Caution Exceeding these electrical parameters decreases the expected life of the
SCXI-1192.
Power Requirement
+5 VDC
Maximum (simultaneous
transition of all relays) .................... 800 mA
Physical
Dimensions............................................. 17.30 by 19.56 by 3.02 cm
(6.81 by 7.70 by 1.19 in.)
I/O connector.......................................... 3 SMA female per relay
8 relays
© National Instruments Corporation
A-9
SCXI-1190/1191/1192 User Manual
Appendix A
Specifications for SCXI-1192
Environment
Operating temperature ......................... 0 to 50 °C
Storage temperature............................. –20 to 70 °C
Relative humidity ...................................10% to 90% noncondensing
Electromagnetic Compatibility
EMC/EMI ...............................................CE, C-Tick, and FCC Part 15
(Class A) Compliant
Electrical emissions ................................EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
Electrical immunity ................................Evaluated to EN 61326-1:1997
A1:1998, Table 1
Note For full EMC and EMI compliance, you must operate this device with shielded
cabling. See the Declaration of Conformity (DoC) for this product for any additional
regulatory compliance information. To obtain the DoC for this product, click
the DoCs by product family. Select the appropriate product family, followed by your
product, and a link to the DoC (in Adobe Acrobat format) appears. Click the Acrobat icon
to download or read the DoC.
Safety
Designed in accordance with IEC 61010-1, UL 3111-1, and CAN/CSA
C22.2 No. 1010.1 for electrical measuring and test equipment.
Installation category II1
Pollution degree 2
1
Category II refers to local-level power distribution, such as that provided by a standard wall outlet.
SCXI-1190/1191/1192 User Manual
A-10
ni.com
B
Common Questions
This appendix addresses common questions you may have while using
your SCXI-119X switching module.
The SCXI-1190 is listed to have a 1.3 GHz bandwidth. Can I use it at
2 GHz bandwidth?
The SCXI-1190 can definitely be used at 2 GHz, as long as the isolation,
VSWR, and attenuation levels are acceptable for your particular
application. Refer to Appendix A, Specifications, for a complete list of the
parameters at 2 GHz.
What should I do if the software detects the module but the switches do
not switch?
•
•
•
Verify that the switches do not switch by trying to close and open the
switches. You can verify the switches functionality by using the
NI-SWITCH Soft Front Panel.
Also be sure to check the return codes of the NI-SWITCH operation to
ensure that there are no errors (negative value) or warnings (positive
value).
Finally, verify that your code is correct. For reference, see the
examples described in the NI-SWITCH Software User Manual.
The NI-SWITCH driver ships with several examples in source code.
Compare your algorithm to those in the examples.
Do I need to program the switch module myself?
The SCXI-119X comes with the NI-SWITCH driver software, which
exports the full functionality of the module. NI-SWITCH handles the
complex issues of direct memory access, interrupts, and operating system
interfacing for you. All you need to do is make high-level calls to these
functions from your ADE.
One of the relays may have worn out. Relays are mechanical devices and
have a limited switching lifetime. Call National Instruments for
information on replacing your relays.
© National Instruments Corporation
B-1
SCXI-1190/1191/1192 User Manual
Appendix B
Common Questions
What is VSWR, and how does it affect my system?
VSWR stands for Voltage Standing Wave Ratio; it is a measure of the
reflection on a transmission line. The percentage of a signal reflected back
to the source is equal to (VSWR-1)/(VSWR+1). This means that a VSWR
of 1 is optimum, with no reflected signal. On higher power applications, a
means that a large percentage of the power from the source cannot be
delivered to the load.
What is isolation, and why is it important?
Isolation is a measure of the amount of crosstalk between channels in a
bank. For example, if a 1 Vp-p signal is isolated by 40 dB to another
channel, then 10 mVp-p will be seen on the adjacent channel (in addition
to the signal already on the channel). Higher channel-to-channel isolation
results in lower crosstalk.
SCXI-1190/1191/1192 User Manual
B-2
ni.com
C
Technical Support Resources
Web Support
National Instruments Web support is your first stop for help in solving
installation, configuration, and application problems and questions. Online
problem-solving and diagnostic resources include frequently asked
questions, knowledge bases, product-specific troubleshooting wizards,
manuals, drivers, software updates, and more. Web support is available
through the Technical Support section of ni.com
NI Developer Zone
The NI Developer Zone at ni.com/zone is the essential resource for
building measurement and automation systems. At the NI Developer Zone,
you can easily access the latest example programs, system configurators,
tutorials, technical news, as well as a community of developers ready to
share their own techniques.
Customer Education
National Instruments provides a number of alternatives to satisfy your
training needs, from self-paced tutorials, videos, and interactive CDs to
instructor-led hands-on courses at locations around the world. Visit the
Customer Education section of ni.com for online course schedules,
syllabi, training centers, and class registration.
System Integration
If you have time constraints, limited in-house technical resources, or other
dilemmas, you may prefer to employ consulting or system integration
services. You can rely on the expertise available through our worldwide
network of Alliance Program members. To find out more about our
Alliance system integration solutions, visit the System Integration section
of ni.com
© National Instruments Corporation
C-1
SCXI-1190/1191/1192 User Manual
Appendix C
Technical Support Resources
Worldwide Support
National Instruments has offices located around the world to help address
your support needs. You can access our branch office Web sites from the
Worldwide Offices section of ni.com. Branch office Web sites provide
up-to-date contact information, support phone numbers, e-mail addresses,
and current events.
If you have searched the technical support resources on our Web site and
still cannot find the answers you need, contact your local office or National
Instruments corporate. Phone numbers for our worldwide offices are listed
at the front of this manual.
SCXI-1190/1191/1192 User Manual
C-2
ni.com
Glossary
Prefix
n-
Meanings
nano-
Value
10–9
10–3
106
m-
milli-
M-
mega-
giga-
G-
109
Symbols
°
degrees
ohms
Ω
%
percent
A
A
amperes
AC
ADC
ADE
alternating current
amperes, direct current
application development environment such as LabVIEW,
LabWindows/CVI, Visual Basic, and C or C++
Arms
amperes, root-mean-square
C
C
Celsius
channel
pin or wire lead on the switch to which you apply or from which you read
the analog or digital signal. Signals can be single-ended or differential.
© National Instruments Corporation
G-1
SCXI-1190/1191/1192 User Manual
Glossary
D
DAQ
data acquisition
decibel
dB
DC
direct current
device
a plug-in board, card, or pad that can contain multiple channels and
conversion devices. Some examples of devices are computers,
multimeters, multiplexers, oscillators, operator interfaces, and
counters.
digital multimeter
a multifunction meter used to make measurements such as voltage, current,
resistance frequency, temperature, and so on
DMM
See digital multimeter
drivers/driver software
software that controls a specific hardware device such as a switch card
H
Hz
hertz—the number of scans read or updates written per second
I
isolation
a measure of the amount of cross-talk between channels
Interchangeable Virtual Instrument
inches
IVI
in.
M
MB
megabytes of memory
multiplexer
a switching device with multiple inputs that sequentially connects each
of its inputs to its output, typically at high speeds, in order to measure
several signals with a single analog input channel
SCXI-1190/1191/1192 User Manual
G-2
ni.com
Glossary
N
NI-SWITCH
an IVI-based instrument driver that supports the National Instruments line
of switch cards
P
PXI
PCI with eXtensions for Instrumentation
radio frequency
R
RF
relay
a switch that connects or disconnects the signal to a common through the
physical movement of a metal arm
rms
root mean square—the square root of the average value of the square of the
instantaneous signal amplitude; a measure of signal amplitude
S
s
seconds
SCXI
Signal Conditioning eXtensions for Instrumentation—the National
Instruments product line for conditioning low-level signals within an
external chassis near sensors so only high-level signals are sent to DAQ
boards in the noisy PC environment
V
V
volts
VDC
Vp-P
volts, direct current
volts peak to peak
© National Instruments Corporation
G-3
SCXI-1190/1191/1192 User Manual
Glossary
Vrms
volts, root-mean-square
VSWR
Voltage Standing Wave Ratio
W
W
watts
SCXI-1190/1191/1192 User Manual
G-4
ni.com
Index
A
D
auto-detection of modules, 1-9 to 1-10
DAQ device. See connecting SCXI-119X to
DAQ device.
dynamic characteristics
SCXI-1190, A-2 to A-3
SCXI-1191, A-6
B
block diagrams
SCXI-1190/1191, 2-2
SCXI-1192, 2-2
SCXI-1192, A-9
E
C
electromagnetic compatibility
SCXI-1190, A-3
certifications and compliances
SCXI-1190, A-4
SCXI-1191, A-7
SCXI-1191, A-7
SCXI-1192, A-10
common questions, B-1 to B-2
configuration, 1-9 to 1-12
environment specifications
SCXI-1190, A-3
auto-detecting modules, 1-9 to 1-10
manually adding modules, 1-10
procedure for, 1-9
SCXI-1191, A-7
SCXI-1192, A-10
expanding RF multiplexer, 2-8 to 2-9
self-test verification, 1-11
troubleshooting self-test
verification, 1-11 to 1-12
F
connecting SCXI-119X to DAQ device
DAQ device or computer-based
instrument, 1-6 to 1-8
in PXI combination chassis, 1-8
connecting signals, 2-5 to 2-9
expanding RF multiplexer, 2-8 to 2-9
multiplexing to RF sources, 2-6 to 2-7
SMA and SMB connectors (figure), 2-5
contact protection, 2-4
functional overview, 2-1 to 2-3
H
hardware installation. See installation.
I
input characteristics
SCXI-1190, A-1
SCXI-1191, A-5
SCXI-1192, A-8
conventions used in manual, vi
customer education, C-1
© National Instruments Corporation
I-1
SCXI-1190/1191/1192 User Manual
Index
installation, 1-2 to 1-8
power requirement specifications
SCXI-1190, A-3
connecting SCXI-119X to DAQ device
DAQ device or computer-based
instrument, 1-6 to 1-8
SCXI-1191, A-6
SCXI-1192, A-9
in PXI combination chassis, 1-8
removing SCXI-119X, 1-13 to 1-14
from Measurement & Automation
Explorer, 1-14
from SCXI chassis, 1-13 to 1-14
into SCXI chassis, 1-4 to 1-6
software installation, 1-2 to 1-3
unpacking, 1-4
PXI combination chassis, connecting with
SCXI-119X, 1-8
Q
questions and answers, B-1 to B-2
R
isolation, defined, B-2
reflective switch, 2-3
removing SCXI-119X, 1-13 to 1-14
from Measurement & Automation
Explorer, 1-14
from SCXI chassis, 1-13 to 1-14
RF multiplexer, expanding, 2-8 to 2-9
RF performance characteristics
SCXI-1190, A-2
M
Measurement & Automation Explorer (MAX)
auto-detecting modules, 1-9 to 1-10
configuration procedure, 1-9
manually adding modules, 1-10
removing SCXI-119X, 1-14
SCXI-1191, A-5 to A-6
self-test verification, 1-11
SCXI-1192, A-8 to A-9
RF sources
troubleshooting, 1-11 to 1-12
multiplexing
multiplexing to, 2-6 to 2-7
RF switching precautions, 2-3 to 2-4
expanding RF multiplexer, 2-8 to 2-9
multiplexing to RF sources, 2-6 to 2-7
S
safety
N
NI Developer Zone, C-1
NI-SWITCH software, 1-2
contact protection, 2-4
RF switching precautions, 2-3 to 2-4
safety instructions, 1-3
safety specifications
P
physical specifications
SCXI-1190, A-3
SCXI-1191, A-6
SCXI-1192, A-9
SCXI-1190, A-4
SCXI-1191, A-7
SCXI-1192, A-10
SCXI chassis
installing SCXI-119X into, 1-4 to 1-6
removing SCXI-119X from, 1-13 to 1-14
SCXI-1190/1191/1192 User Manual
I-2
ni.com
Index
SCXI-119X
block diagrams, 2-2 to 2-3
electromagnetic compatibility, A-7
environment, A-7
input characteristics, A-5
physical, A-6
power requirement, A-6
RF performance characteristics,
A-5 to A-6
common questions, B-1 to B-2
configuration and self-test, 1-9 to 1-12
connecting signals, 2-5 to 2-9
contact protection, 2-4
functional overview, 2-1 to 2-3
installation, 1-2 to 1-8
safety, A-7
SCXI-1192, A-8 to A-10
dynamic characteristics, A-9
electromagnetic compatibility, A-10
environment, A-10
overview, 1-1
RF switching precautions, 2-3 to 2-4
safety instructions, 1-3
specifications
input characteristics, A-8
physical, A-9
SCXI-1190, A-1 to A-4
SCXI-1191, A-5 to A-7
power requirements, A-9
RF performance characteristics,
A-8 to A-9
SCXI-1192, A-8 to A-10
self-test verification
procedure for, 1-11
safety, A-10
system integration, by National Instruments,
C-1
troubleshooting, 1-11 to 1-12
signal connections, 2-5 to 2-9
expanding RF multiplexer, 2-8 to 2-9
multiplexing to RF sources, 2-6 to 2-7
SMA and SMB connectors (figure), 2-5
software installation, 1-2 to 1-3
specifications
T
technical support resources, C-1 to C-2
troubleshooting common problems,
B-1 to B-2
SCXI-1190, A-1 to A-4
certifications and compliances, A-4
dynamic characteristics, A-2 to A-3
electromagnetic compatibility, A-3
environment, A-3
input characteristics, A-1
physical, A-3
V
Voltage Standing Wave Ratio (VSWR), B-2
power requirement, A-3
RF performance characteristics, A-2
safety, A-4
W
Web support from National Instruments, C-1
Worldwide technical support, C-2
SCXI-1191, A-5 to A-7
certifications and compliances, A-7
dynamic characteristics, A-6
© National Instruments Corporation
I-3
SCXI-1190/1191/1192 User Manual
|